Concentric Vane Repulsion Type Instrument

Concentric Vane Repulsion Type Instrument (Moving Iron Type):

Concentric Vane Repulsion Type Instrument – A variation of the radial vane instrument is the concentric vane repulsion movement. The instrument has two concentric vanes.

One vane is rigidly attached to the coil frame while the other can rotate coaxially inside the stationary vane, as shown in Fig. 2.9. Both vanes are magnetised by the current in the coil to the same polarity, causing the vanes to slip laterally under repulsion. Because the moving vane is attached to a pivoted shaft, this repulsion results in a rotational force that is a function of the current in the coil. As in other mechanisms the final pointer position is a measure of the coil current. Since this movement, like all iron vane instruments, does not distinguish polarity, the concentric vane may be used on dc and ac, but it is most commonly used for the latter.

Concentric Vane Repulsion Type Instrument

Damping is obtained by a light aluminium damping vane, rotating with small clearance in a dosed air chamber. When used on ac, the actual operating torque is pulsating and this may cause vibration of the pointer. Rigid (trussed) pointer construction effectively eliminates such vibration and prevents bending of the pointer on heavy overloads. The concentric vane moving iron instrument is only moderately sensitive and has square law scale characteristics. The accu­racy of the instrument is limited by several factors: (i) the magnetisation curve of the iron vane is non-linear. (ii) at low current values, the peak to peak of the ac produces a greater displacement per unit current than the average value, re­sulting in an ac reading that may be appreciably higher than the equivalent dc reading at the lower end of the scale. Similarly, at the higher end of the scale, the knee of the magnetisation curve is approached and the peak value of the ac produces less deflection per unit current than the average value, so that the ac reading is lower than the equivalent dc value.

(Hysteresis in iron and eddy currents in the vanes and other metal parts of the instrument further affect the accuracy of the reading.) The flux density is very small even at full scale values of current, so that the instrument has a low cur­rent sensitivity. There are no current carrying parts in the moving system, hence the iron vane meter is extremely rugged and reliable. It is not easily damaged even under severe overload conditions.

Adding a suitable multiplier converts the iron vane movement into a voltme­ter; adding a shunt produces different current ranges. When an iron vane move­ment is used as an ac voltmeter, the frequency increases the impedance of the instrument and therefore a lower reading is obtained for a given applied volt­age. An iron vane voltmeter should therefore always be calibrated at the fre­quency at which it is to be used. The usual commercial instrument may be used within its accuracy tolerance from 25-125 Hz.